processconvert
Instrumentation

Instrument Range vs Span

The range is the pair of endpoints LRV and URV; the span is the width URV − LRV. Learn why a range can cross zero, how elevated and suppressed zeros work, and how range and span drive signal scaling.

TypeEngineering guide — concept explainer

Definition

An instrument's range is the pair of endpoints it is calibrated between: the lower range value (LRV) and the upper range value (URV), written as LRV to URV. The span is a single number — the width of that range, span = URV − LRV. A 0–100 kPa instrument has a range of 0 to 100 kPa and a span of 100 kPa; a −50 to 150 °C instrument has a range crossing zero and a span of 200 °C.

Why it matters

Range and span are the two numbers that define how an instrument maps the physical world onto its signal, and confusing them is a classic source of error. The span sets the resolution and the percent-of-span scaling; the range endpoints set where zero sits and whether the measurement can go negative. Ranging a transmitter, choosing turndown, sizing for the expected operating window, and reading a calibration sheet all depend on keeping range and span distinct. Many process ranges deliberately cross zero — compound pressure from vacuum to positive, or temperatures spanning freezing — which only makes sense once you separate the endpoints (range) from the width (span).

Formula

Span
span = URV − LRV
Percent of span
% span = (PV − LRV) / span × 100
Process value from percent
PV = LRV + span × % span / 100

Units involved

  • LRV — lower range value, the process value at 0% of span (4 mA)
  • URV — upper range value, the process value at 100% of span (20 mA)
  • range — the endpoint pair, LRV to URV, in engineering units
  • span — URV − LRV, the width of the range, in the same units
  • PV — process value in engineering units
  • % span — position of PV within the range, 0% to 100%

Concept diagram

0 to 100 kPa · span 100LRV 0URV 100−50 to 150 °C · span 200 (crosses zero)0 °CLRV −50URV 150

Worked example

Compare two instruments: a 0–100 kPa pressure transmitter and a −50 to 150 °C temperature transmitter. Find the span of each, and the percent of span at a process value of 50.

  1. 01Pressure: span = 100 − 0 = 100 kPa; at PV = 50 kPa, % span = (50 − 0) / 100 × 100 = 50%
  2. 02Temperature: span = 150 − (−50) = 200 °C
  3. 03At PV = 50 °C, % span = (50 − (−50)) / 200 × 100
  4. 04% span = 100 / 200 × 100 = 50%
Result

Both read 50% of span at PV = 50, but the spans differ (100 vs 200) and the °C range crosses zero with an elevated zero at LRV = −50.

Common mistakes

  • Using the URV as the span. Span is URV − LRV, not URV. A 10–50 bar instrument has URV = 50 but span = 40.
  • Assuming the LRV is zero. A suppressed zero (LRV > 0, e.g. 50–250 °C) and an elevated zero (LRV < 0, e.g. −50 to 150 °C) are both common. Percent of span is always measured from the LRV.
  • Thinking a range cannot be negative. Many ranges cross zero — compound pressure or sub-zero temperature. The span is still URV − LRV and stays positive as long as URV > LRV.
  • Comparing instruments by range when you mean resolution. Two instruments can share a span but sit at different ranges, or share a range endpoint but have different spans. Be explicit about which you are comparing.

When to use the calculator

Use the Instrument Range / Span calculator to compute the span from LRV and URV, find the process value at a given percent of span, or find the percent of span for a process value — including ranges that cross zero. Pair it with the mA to Process Value and Process Value to mA calculators to see how the same range maps onto the 4–20 mA signal.

FAQ

What is the difference between range and span in one sentence?
The range is where the instrument measures (LRV to URV); the span is how wide that window is (URV − LRV).
Why would a range cross zero?
Because the process does. Compound pressure instruments read from vacuum (negative gauge) to positive pressure, and temperature instruments often span below and above 0 °C. Crossing zero is normal and just means the LRV is negative.
What are suppressed and elevated zeros?
A suppressed zero means the LRV is above zero (e.g. 50–250 °C, so 0% of span is 50 °C). An elevated zero means the LRV is below zero (e.g. −50 to 150 °C). Both are handled by the same formula, percent of span measured from the LRV.
How do range and span affect the 4–20 mA signal?
The signal always runs 4 mA at the LRV to 20 mA at the URV, regardless of the actual range. The span determines how many engineering units each milliamp represents: a narrow span gives finer resolution per mA, a wide span gives coarser resolution.

Related calculators

Related conversions

Related guides